Thông tin tài liệu

Thông tin siêu dữ liệu biểu ghi
Trường DC Giá trịNgôn ngữ
dc.contributor.authorAna C., Gimenez-Ingalaturre-
dc.contributor.authorKhaoula, Ben-Jeddou-
dc.contributor.authorJosefina, Perez-Arantegui-
dc.date.accessioned2023-04-20T02:18:35Z-
dc.date.available2023-04-20T02:18:35Z-
dc.date.issued2022-
dc.identifier.urihttps://link.springer.com/article/10.1007/s00216-022-04215-z-
dc.identifier.urihttps://dlib.phenikaa-uni.edu.vn/handle/PNK/8114-
dc.descriptionCC BYvi
dc.description.abstractSingle particle inductively coupled plasma mass spectrometry (SP-ICP-MS) is a technique widely used to obtain direct information about the number concentration and the size distribution of nanoparticles in liquid suspensions. However, its methods still lack clear quality control strategies to confirm the validity of the information derived from them. Only the detection of the complete size distribution of the nanoparticles in a sample over the size critical value ensures obtaining unbiased quantitative information, otherwise information should be restricted to report the presence of nanoparticles over a certain size and number concentration since their actual total number concentration is underestimated and the size overestimated.vi
dc.language.isoenvi
dc.publisherSpringervi
dc.subjectSP-ICP-MSvi
dc.titleHow to trust size distributions obtained by single particle inductively coupled plasma mass spectrometry analysisvi
dc.typeBookvi
Bộ sưu tập
OER - Khoa học Tự nhiên

Danh sách tệp tin đính kèm: