Item Infomation
Full metadata record
DC Field | Value | Language |
---|---|---|
dc.contributor.author | Lidia, Fijałkowska-Lichwa | - |
dc.contributor.author | Tadeusz A., Przylibski | - |
dc.contributor.author | Maciej, Norenberg | - |
dc.date.accessioned | 2023-04-21T08:10:46Z | - |
dc.date.available | 2023-04-21T08:10:46Z | - |
dc.date.issued | 2023 | - |
dc.identifier.uri | https://link.springer.com/article/10.1007/s10967-023-08882-9 | - |
dc.identifier.uri | https://dlib.phenikaa-uni.edu.vn/handle/PNK/8219 | - |
dc.description | CC BY | vi |
dc.description.abstract | The semi-conductor SRDN-3a probe, an AlphaE detector with a silicon diode in a diffusion chamber, an AlphaGUARD monitor with an ionization chamber and CR-39 track detectors had been tested simultaneously in long-term measurements conducted in the technical corridor of Dobromierz dam. The passive detectors were exposed twice: for 56 days, and 117 days, others in parallel, with a 1-h data recording interval. The data distribution was tested with the Shapiro–Wilk test and outliers in the critical region were identified using Shewhart control charts. The correctness was evaluated by a z-score test recommended by the IAEA. The characteristics of outliers for each detectors are determined by the location of the critical region (a two-sided region and a positive skew). | vi |
dc.language.iso | vi | vi |
dc.publisher | Springer | vi |
dc.subject | SRDN-3a probe | vi |
dc.subject | AlphaE detector | vi |
dc.title | Intercomparison of equipment measuring radon activity concentration in the air—an example from a hydrotechnical structure in Dobromierz (SW Poland) | vi |
Appears in Collections | ||
OER - Khoa học Tự nhiên |
Files in This Item: