Item Infomation
Title: | Intercomparison of equipment measuring radon activity concentration in the air—an example from a hydrotechnical structure in Dobromierz (SW Poland) |
Authors: | Lidia, Fijałkowska-Lichwa Tadeusz A., Przylibski Maciej, Norenberg |
Issue Date: | 2023 |
Publisher: | Springer |
Abstract: | The semi-conductor SRDN-3a probe, an AlphaE detector with a silicon diode in a diffusion chamber, an AlphaGUARD monitor with an ionization chamber and CR-39 track detectors had been tested simultaneously in long-term measurements conducted in the technical corridor of Dobromierz dam. The passive detectors were exposed twice: for 56 days, and 117 days, others in parallel, with a 1-h data recording interval. The data distribution was tested with the Shapiro–Wilk test and outliers in the critical region were identified using Shewhart control charts. The correctness was evaluated by a z-score test recommended by the IAEA. The characteristics of outliers for each detectors are determined by the location of the critical region (a two-sided region and a positive skew). |
Description: | CC BY |
URI: | https://link.springer.com/article/10.1007/s10967-023-08882-9 https://dlib.phenikaa-uni.edu.vn/handle/PNK/8219 |
Appears in Collections | OER - Khoa học Tự nhiên |
ABSTRACTS VIEWS
22
FULLTEXT VIEWS
4
Files in This Item: