Item Infomation


Title: Intercomparison of equipment measuring radon activity concentration in the air—an example from a hydrotechnical structure in Dobromierz (SW Poland)
Authors: Lidia, Fijałkowska-Lichwa
Tadeusz A., Przylibski
Maciej, Norenberg
Issue Date: 2023
Publisher: Springer
Abstract: The semi-conductor SRDN-3a probe, an AlphaE detector with a silicon diode in a diffusion chamber, an AlphaGUARD monitor with an ionization chamber and CR-39 track detectors had been tested simultaneously in long-term measurements conducted in the technical corridor of Dobromierz dam. The passive detectors were exposed twice: for 56 days, and 117 days, others in parallel, with a 1-h data recording interval. The data distribution was tested with the Shapiro–Wilk test and outliers in the critical region were identified using Shewhart control charts. The correctness was evaluated by a z-score test recommended by the IAEA. The characteristics of outliers for each detectors are determined by the location of the critical region (a two-sided region and a positive skew).
Description: CC BY
URI: https://link.springer.com/article/10.1007/s10967-023-08882-9
https://dlib.phenikaa-uni.edu.vn/handle/PNK/8219
Appears in CollectionsOER - Khoa học Tự nhiên
ABSTRACTS VIEWS

22

FULLTEXT VIEWS

4

Files in This Item: